Proservo NMS5
72
Hauser
G8V4H1 OPE. POINT
G8V4H4 BAL. LEVEL
G8V4H5 UP WAIT TIME
Item
Procedure
Remarks
Figure. 70:
G8V4H1
1.
Select G8V4H1 "OPE
POINT " to determine
number of density
measuring points.
•
Set access code to 51.
Default value is 2 points
and selectable up to 16
points.
2.
Enter required number
by pr or - key,
and select E to enter.
•
Whatever the number of
measuring point is
selected, the measuring
interval & point between
each point are automati-
cally calculated within the
liquid level.
Item
Procedure
Remarks
Figure. 71:
G8V4H4
1.
Select G8V4H4 "BAL
LEVEL" to determine
the liquid stability range
in order to start the
operation.
•
Set access code to 51.
NMS software has a func-
tion to scan & record the
liquid level every 5 min-
utes.
2.
Enter desired value by
pr or - key and
select E to enter. Input
value of 99.9mm
represents that the
measuring procedure
starts regardless of level
condition.
•
NMS5 only starts the
actual measurement when
the liquid level is within
the entered value
compared to the recorded
level data 5 minutes ago.
This function ensures the
safe operating condition to
avoid filling or draining
liquid from the tank.
Item
Procedure
Remarks
Figure. 72:
G8V4H5
1.
Select G8V4H4 "UP
WAIT TIME" to
determine the displacer
weight calibration
interval in air.
•
Set access code to 51. The
displacer raises approxi-
mately 500mm above liq-
uid surface after confirm-
ing the stability of liquid
level within "BAL LEVEL"
function.
2.
Enter desired waiting
time by pr or -
key, and select E to
enter.
•
The displacer weight is
calibrated at 8 different
positions in air and the
interval between those
positions are adjustable
within this function to
allow excess liquid to drip
off the displacer to achieve
higher accuracy.
PROFILE
OPE.
OPE.
POINT
Function
Group
Matrix Group: TANK PROFILE
TANK
PROFILE
Function
Group
BAL.
LEVEL
GVH=844
Matrix Group : TANK PROFILE
TANK
PROFILE
Function
Group
UP.WAIT
TIME
GVH=845
Matrix Group : TANK PROFILE
7 Commissioning