Proservo NMS5
Hauser
121
50
Matrix group
Function group
Item
Access
code
Short description
Default value
Set
Select
Display
Possible settings,
selection, or display
Index No,
GVH
CALIBRATION
LEVEL DATA
TANK HEIGHT
Elevation level of manual dipping reference.
0 - 99999.9 mm
140
DISPL. RAISE DENS
144
16000.0 mm
Set
0 - 300 mm
143
150 mm
DISPL. SUBM. DENS
LEVEL BELOW 0
149
DIP POINT OFFSET
50
Difference between Tank Height and Reference Position of
Proservo NMS5. This data is automatically adjusted by
Proservo software when Set Level procedure is performed.
0.0 mm
Set
0 - 99999.9 mm
141
DISPLAC. DRAFT
50
Set the draft of displacer. This value is used to calculate G0V0H4
Bottom Level measurement.
10.0 mm
Set
0 - 999.9 mm
142
Diameter
3PECIlCATION
Draft
30 mm
40 mm
50 mm
50 mm
70 mm
110 mm
Special
Special
Std
Conical
NMi
PTB
Varies
Varies
45 mm
35 mm
16.9 mm
10 mm
Typical Displacer Draft Settings
Set distance for displacer to rise above level position during
Density operations.
Observe the rule: G1V4H4 = 300 n, where = integer
51
Set distance for displacer to sink below level position during
Density operations.
Observe the rule: G1V4H4 = 300 n, where = integer
51
Select method for handling negative level data in V1 protocol
communication. Select "99999 mm" to display negative level data
backwards from 99999 mm. Select "0 mm" to display all negative
level data as 0 mm.
150 mm
99999 mm
0 - 1500 mm
0 - 99999 mm
Select
Set
Set
51
13 Matrix