Do not let wire fall off the wire drum.
4.
Lift the displacer until Sensor Count is stable and record the value of Sensor Count (Wa1).
•
|
Wa2 - Wa1
|
= 10000 (+/-100)?
5.
If Yes, go to step 7 "GVH740 A Zero".
6.
If No, return the displacer to the original position at GVH741
•
If
|
Wa2 - Wa1
|
<10,000, increase Index Count.
•
If
|
Wa2 - Wa1
|
>10,000, decrease Index Count.
•
Repeat until
|
Wa2 - Wa1
|
= 10000 (+/-100).
7.
Adjust A Zero Sensor Count = 21000 (+/-100) at GVH740.
Figure. 93:
Hall Sensor Adjustment 3
8.
Record the value of Sensor Counter (Wa2)
9.
Lift the displacer until Sensor Count is stable and record the value of Sensor Count (Wa2).
•
|
Wa1 - Wa2
|
= 10000 (+/-100)?
10. If yes, go to step 12. "GVH743 B Span".
11. If no, repeat Steps 3and 7.
12. Set B Span at GVH743 and repeat steps above.
Figure. 94: Hall Sensor Adjustment 4
13. Adjust B Zero Sensor Count = 11000 (+/- 100) at GVH742.
Figure. 95:
Hall Sensor Adjustment 5
This completes the hall sensor adjustment.
29XX 21000
ADJ. A ZERO
(Ex Wa2=2100)
(Ex Wa1=1100)
29XX 11000
ADJ. A ZERO
31XX 11000
ADJ. B SPAN
31XX 21000
ADJ. B SPAN
(Ex Wa1=1100)
31XX 21000
ADJ. B SPAN