Proservo NMS5
104
Hauser
Figure. 103: Calibration 3-B
5. Replace the displacer
with 50 x g weight and
wait until Sa and Sb are
stable.
6. Press "E" and (-) key
together.
Figure. 104: Calibration 4-B
7. Remove the 50xg wight
from the displace and
mount the displacer on
the wire.
8. Press "E" and (+) keys
together at "Displacer
on: E & "+".
•
Calibration starts
automatically, which
takes approximately 10
minutes.
Figure. 105: Calibration 5-B
9. Check /-
2.0 grams? If yes, cali-
bration is done.
If no, make sure there is
no vibration during cali-
bration
10.Repeat Hall sensor
adjustment and calibra-
tion.
Item
Procedure
Remarks
11 Displacer and Measuring Wire