Proservo NMS5
Hauser
133
Matrix group
Function group
Item
Access
code
Short description
Default value
Set
Select
Display
Possible settings,
selection, or display
Index No,
GVH
SERVICE
MEASURED WEIGHT
370
372
371
51
51
SERVICE
WEIGHT CALIBR.
RELE. OVER TENS
DRUM SETTING
51
373
Display
Displays tension on measuring wire as calculated by NMS CPU
module.
0.0 g
Select
Release displacer overtension condition.
Note : First, set G0V2H0. Operation = STOP.
Aligns wire drum with internal detector unit. When displacer is
shipped separately, and wire drum is removed to install displacer,
set this function prior to making new Weight Calibration.
Not required for All-in-One shipments of NMS5 Proservo.
Initiates Weight Table recalibration procedure. Overwrites existing
weight table with new weight table.
Caution!: Once recalibration procedure begins it must be
completed, i.e. it connot be stopped or reversed.
OFF
ON
OFF
ON
Select
Select
OFF
OFF
OFF
0.0 - 999 g
OFF
ON
SERVICE VALUE
SENSOR VALUE
380
51
Displays the A/D values from the encoder.
Sa 21000:
A 21000
Sb 11000:
B 11000
WT. COUNT CAL A
395
WT. COUNT CAL. B
51
394
51
Displays A/D and displacer weight correction values for Sensor A
Use the + and - key to scroll through 50 points.
Displays A/D and displacer weight correction values for Sensor B
Use the + and - key to scroll through 50 points.
0 0 0.0 g
0 0 0..0 g
Display
Display
SERVICE DATA
~ ~ ~ ~ ~ ~ ~ ~
DISPL. REFERENCE
ZERO ADJ. WEIGHT
51
51
Length for the starting position of weight calibration.
The displacer will stop at this position without regard to the high
stop level when it is hoisted, if the high stop level is set higher than
this point.
Low weight for weight calibration.
70 mm
Set
10...999 mm
378
0.0 g
Set
0.0...999.9 g
379
Display
13 Matrix