Proservo NMS5
58
Hauser
7.3
Settings for Connection to NRF560
To connect NRF560, NMS5 requires the following settings.
Caution!
Turn on the power to NMS5 first.
Item
Procedure
Remarks
Figure. 61:
Matrix Group: Service
1.
In Static Matrix
"MORE
FUNCTION", invoke
GVH=030 "MATRIX
OF" and select
"SERVICE".
2.
Invoke the Dynamic
Matrix GVH=361
"CONNECTION
NRF" screen.
3.
Use the "+" and "-"
keys to select either
"CONTACT 1" or
"CONTACT 2".
•
Set access code 51.
•
CONTACT 1...
NRF 560
software version 1.6x
and earlier (those NRF
560 that indicate no
software version
correspond to
connection type 1).
•
CONTACT 2...
NRF 560
software version 1.8x
and later.
MORE
FUNCTION
SYSTEM
DATA
CONNECTION NRF
3 6 1
MATRIX OF
0 3 0
7 Commissioning