Proservo NMS5
Hauser
103
Weight Table Calibration and Density
Item
Procedure
Remarks
Figure. 101: Calibration 1-B
1. Set Access Code at 51
and confirm the
following data.
•
G3V4H0 wire Drum
Circum.(engraved on
wire drum)
•
G3V4H1 Wire
Weight Standard
SUS = 1.40
PFA=4.55
Hastelloy=2.48
•
G3V4H2 Displacer
Weight (engraved on
displacer)
•
G3V7H9 Zero Adjust
Weght = e.g. 50.2
2. Confirm each engraved
data in the wire drum
and the displacer and
values in Matrix are the
same.
Figure. 102: Calibration 2-B
3. Displacer is located in
the calibration window
or the maintenance
chamber.
Set Operation = STOP
Set G3V7H9=e.g.
50.2g (test weight, in
grams)
Set G3V7H3 Weight
Calib. =ON
4. Enter No "-" key at
"Low weight set?" cali-
bration window and
enter Yes "+" key for
maintenance
chamber.
e.g.
300.110
e.g.
250.2 g
H
V
0
1
2
4
xxx.xxx
x.xx
xxx.x
Service
Matrix: G3
Wire Drum
Displacer
51
ACCESS CODE
H
V
7
9
50.2
50.x g
weight
11 Displacer and Measuring Wire