Proservo NMS5
Hauser
71
•
A. Tank profile for up to 16 points for total liquid height
G8V4H0 Ope Select: Select 1: Tank Profile.
•
B. Tank profile for upper phase liquid only
G8V4H0 Ope Select: Select 2: I/F Profile.
•
C. Tank profile for based on a manually entered profile parameter.
G8V4H0 Ope Select: Select 3: MANU.I/F Profile.
When choosing A, B, and C above, the following parameters can be set or select. The factory default settings
also can be used.
Initial Setting of Tank Profile (Density Profile) Measurement
All of necessary set up can be done at "G8 Tank Profile" matrix. There are six factors to configured prior to the
actual operation.
G8V4H0 OPE. SELECT
Matrix
Item
Setting
G8V4H1
Ope Point
Set the number of measurement points (2-16) for Tank Profile mea-
surement.
Default setting = 2
.
G8V4H3
I/F Manu.
Level
Manually set the water interface level (0-99999.9mm) to be
referenced during the tank profile measurement.
Default setting = 0
mm
.
Only used with C, tank profile for based on a manually entered
profile parameter
.
G8V4H4
Bal. Level
Set the allowance for level movement (1.0 -99.9 mm) prior to Tank
Profile operation. If the level movement exceeds this value, profile
operation is canceled. Setting 99.9mm allows Profile measurement to
proceed, regardless of level movement.
Default setting = 2.0 mm
.
G8V4H5
Up Wait Time
Set waiting time (1-31 minutes) for displacer weight to be measured in
air, when making the weight table at start of profile operation. Default
setting = 1 minute.
G8V4H6
Liq Wait Time
Set the time (1-31 minutes) for displacer to stop at each measurement
position during profile operation.
Default setting = 1 minute
.
G8V4H7
Ope Wait Time
Set maximum waiting time (1-31 minutes) for conditions in G8V4H4
(Bal Level) to be satisfied before profile operation begins. E.g. if the
level movement exceeds setting at G8V4H4, NMS5 is in "standby" for
the time set in this matrix. After the maximum time elapses, profile
operation is canceled.
Default setting = 1 minute
.
Item
Procedure
Remarks
Figure. 69:
G8V4H0
1.
Select G0V3H0
"MATRIX OF" in the
row "MORE
FUNCTION" on Static
Matrix, then select
"Tank profile".
•
Set access code to 50.
2.
Select function group
"PROFILE OPE" and
G8V4H0 "OPE
SELECT".
•
The row of PROFILE OPE
is the only function group
that needs configuration
for the density profile
operation.
3.
Select 1: "TANK
PROFILE" by pr
or - key.
4.
Press E to configure.
•
Selection of "0: SPOT"
enables Upper Density
measurement instead of
Density Profile
measurment.
Static
Matrix
MORE
FUNCTION
Matrix of
TANK
PROFILE
PROFILE
OPE.
OPE.
SELECT
STATUS
DATA
DENSITY
1-10
DENSITY
11-16
Position
1-10
Position
10-16
Dynamic
Matrix
Function
Group
GVH=840
GVH=030
Matrix Group : TANK PROFILE
7 Commissioning