Proservo NMS5
Hauser
75
Density Profile Operation Diagram
Figure. 76:
Density Profile
Operation Procedure
1.
Set NMS5 so that it can measures the liquid level.
2.
Raises the displacer and measure each weight of 8 measuring points in the air.
3.
Measure the density at the setting position in the liquid level.
4.
Return to level measurement.
5.
Save the data in G8 matrix accordingly.
This completes the operation procedure.
Caution!
The average density and temperature data regarding density profile function (G8 Matrix) are completely differ-
ent from the data on G0 static matrix which displays the conventional upper density (spot) and temperature
(continuously scanned) data.
8 points of air weight measurement
(within 300mm circumference
of wire drum)
Nozzle
Proservo NMS5
㻼㼞㼛㼟㼑㼞㼢㼛
First Displacer Weight Calculation Line
(Approximately 500mm from Liquid Level)
No. 1 Measuring Point
No. 2 Measuring Point
No. 3 Measuring Point
No. 4 Measuring Point
No. 5 Measuring Point
No. 6 Measuring Point
No. 7 Measuring Point
If required density measuring points are 7, the method to calculate
each point from liquid level in NMS with ver 4.25 softwear is:
(7 1)
Liquid level
x measuring point number
Ex. if the liquid level is 4,000 mm
and needs to determine the position of No. 5 measuring point from the liquid surface;
(7 +1)
4,000
x 5 = 2,500 mm
)NTERFACE0ROlLEUPPERLAYER
Max.16 point
4ANKPROlLELIQUIDLEVELTOTANKBOTTOM
Max.16 point
Liquid Level
7 Commissioning