Gammapilot M FMG60 PROFIBUS PA
Installation
Hauser
13
3.2.2
Installation conditions for level measurement
Conditions
• For level measurements the Gammapilot M is mounted vertically; if possible the detector
head should point downwards.
• The exit angle of the source container must be exactly aligned to the measuring range of
the Gammapilot M. Observe the measuring range marks of the Gammapilot M.
• In cascading mode no gap should occur between the measuring ranges of the different
Gammapilot M.
• The source container and the Gammapilot M must be mounted as close to the vessel as
possible. Any access to the beam must be blocked so that no persons or part of their body
(hand, arm, head) may come into the area of the beam.
• In order to enlarge the lifetime, the Gammapilot M should be protected against direct sun.
If necessary, a protective cover should be applied.
• The mounting device FHG60 (
ä
81, "Accessories") or an equivalent mounting device
should be used for fastening the Gammapilot M.
The mounting device must be installed in a way such that it can withstand the weight of
the Gammapilot M
1)
under all operating conditions (e.g. vibrations).
Note! Tmmapilot M should be given additional support to prevent damage to the
connecting cable or to the unit if it falls off.
NOTICE
The Gammapilot M should be given additional support to prevent damage to the
connecting cable or to the unit if it falls off.
Examples
A0018074
A
Vertical cylinder; the Gammapilot M is mounted vertically with the detector head pointing downwards;
the gamma ray is aligned to the measuring range.
B
Cascading of multiple Gammapilot M; there is no gap between the measuring ranges
C
Wrong: Gammapilot M mounted inside the tank insulation
D
Conical tank outlet (here with sun protection cover)
E
Horizontal cylinder (here with sun protection cover)
F
Right: Tank insulation removed for Gammapilot M
1
Support
1)
The weights of the various versions of the Gammapilot M are summarized in the section "Dimensions/Weight".
A
D
B
C
E
F
1