Gammapilot M FMG60 PROFIBUS PA
Commissioning
Hauser
59
6.3.10 "Pipe diameter" (*0A)
A0018109
Always specify the complete irradiated measuring path L in the "pipe diameter" (*0A) function. Depending on the installation, this
value may be larger than the actual pipe diameter.
On-site display
Meaning
Pipe diam.
*0A
This function is needed for density and concentration measure-
ments only. It is used to specify the irradiated measuring path
L. With standard installation, this value is identical to the inner
pipe diameter D
I
. For other installations (in order to enlarge
the irradiated measuring path) it may be larger (see figure).
The pipe walls are
not
to be considered a part of the measuring
path.
200 mm
1
1
1
L
L
L
a
D
I
a
L
3
0
°
2
.
0
x D
I