Installation
Gammapilot M FMG60 PROFIBUS PA
14
Hauser
3.2.3
Installation conditions for level limit detection
Conditions
• For level limit detection, the Gammapilot M should be mounted horizontally at the height
of the desired level limit.
• The exit angle of the source container must be exactly aligned to the measuring range of
the Gammapilot M. Observe the measuring range marks of the Gammapilot M.
• The source container and the Gammapilot M must be mounted as close to the vessel as
possible. Any access to the beam must be blocked so that no persons or part of their body
(hand, arm, head) may come into the area of the beam.
• In order to enlarge the lifetime, the Gammapilot M should be protected against direct sun.
If necessary, a protective cover should be applied.
• The mounting device FHG60 (
ä
81, "Accessories") or an equivalent mounting device
should be used for fastening the Gammapilot M.
The mounting device itself must be fitted in a way such that it can withstand the weight of
the Gammapilot M
2)
under all operating conditions to be expected.
Examples
A0018075
A
Maximum fail-safe mode
B
Minimum point level detection
2)
The weights of the various versions of the Gammapilot M are summarized in the section "Dimensions/Weight".
A