Commissioning
Gammapilot M FMG60 PROFIBUS PA
74
Hauser
6.5.3
Calibration points
Excerpt from the operating menu
The following excerpt from the operating menu shows, how the density calibration points
are entered. The individual functions are explained in the sections below.
The functions are only accessible after the background calibration has been performed.
"Calibration point" (*1A)
Further options:
• "1" to "9" : Calibration points for various densities
• "10": recalibration point
After entering of the recalibration point, I
0
is recalculated according to the current
measuring conditions. The absorption coefficient
is kept unchanged from the original
calibration.
The calibration point "10" can be entered if the measuring conditions have changed, e.g.
due to buildup in the measuring tube.
*1A
calibr. point
1...9
point 10 for
recalibration
*15
calibration
0: stop/edit
1: start
*11
avg. pulse
rate
*1B
density calib.
(c/s)
*1C
density value
Group
selection
*1D
calibr. point
0: not used
1: used
2: clear
*1E
absorp.
coeff.
*1F
ref. pulse rate
*19
next point
0: no
1: yes
On-site display
Meaning
calibr. point
*1A
This function is used to select, which calibration point will be
entered.
1
2
3
E
0
1