Commissioning
Gammapilot M FMG60 PROFIBUS PA
54
Hauser
6.2
Switching on the device
NOTICE
Error messages A165 "electronics defect" and A635 "present date not defined"
The Gammapilot M contains 2 real-time clocks for the decay compensation, which are
permanently compared to each other for safety reasons. In order to bridge voltage
interruptions, the clocks are buffered with a capacitor. To ensure that the clocks work
correctly and retain the date in the event of a voltage interruption, this capacitor must have
a minimum charge. If the A165 "Electronics defect" or A635 "Present date not defined" error
message appears
after switching on
the Gammapilot M, then the capacitor may possibly
not yet be charged sufficiently. In this case, the Gammapilot M must be operated at the
operating voltage for at least 20 to 30 minutes, in order to charge the capacitor. After this,
the date must be entered correctly. If the error message still persists subsequent to this, it
can be deleted by switching the Gammapilot M off and on.
After switching on the supply voltage, the instrument is first initialized. Due to internal
memory tests, this takes approx. 2 minutes.
On-site display
Meaning
FMG60
Then, the following appear for approximately five seconds:
• Device type
• Software version
• Type of the communication signal
V01.03.06 PA
On first power up you are requested to select the language for
the display texts.
Language
092
Select the language with the
O
and
S
keys.
Confirm your choice by pressing
F
twice.
English
Français
Español
After that the measured value display appears.
Measured value
000
Press
F
to switch to the group selection.
63.4%
Group selection
The "basic setup" function group is already selected.
Press
F
to enter the first function of this function group.
Basic setup
Calibration
Safety settings