Gammapilot M FMG60 PROFIBUS PA
Commissioning
Hauser
77
"Absorption coefficient" (*1E)
NOTICE
If only one calibration point is currently active, the absorption coefficient is not
calculated. The last valid value is used instead. At the first commissioning or after a
reset, the default value,
= 7.70 mm
2
/g, is used. The value can be changed by the user.
"Reference pulse rate" (*1F)
NOTICE
I
0
is the pulse rate for the tube being empty (theoretical reference value). Generally, the
value is significantly higher than any real pulse rate occurring during the
measurement.
"Next point" (*19)
Options:
•
no
This option must be selected if no further calibration point is to be entered or changed.
After this selection the Gammapilot M returns to the group selection and the calibration
is completed.
•
yes
This option must be selected, if a further calibration point is to be entered or changed. The
Gammapilot M returns to the
"calibr. point" (*1A)
function and the next point can be
entered or changed.
On-site display
Meaning
absorp. coeff.
*1E
This function displays the absorption coefficient which results
from the currently active calibration points. The displayed
value should be used for plausibility checking.
7.70 mm2/g
On-site display
Meaning
ref. pulse rate
*1F
This function displays the reference pulse rate I
O
, which results
from the currently active calibration points. The value cannot
be edited.
31687 cps
On-site display
Meaning
next point
*19
This function is used to specify, if a further calibration point is
to be entered or not.
no
yes