Gammapilot M FMG60 PROFIBUS PA
Troubleshooting
Hauser
89
W642 I_back calibration running
Calibration of the current readback path active
A631
background not calibrated
Perform background calibration
(
"calibration" (*1)
function group)
A632
full/covered not calibrated
Perform full/covered calibration
(
"calibration" (*1)
function group)
A633
empty/free not calibrated
Perform empty/free calibration
(
"calibration" (*1)
function group)
A634
density not calibrated
• Check: Has at least one calibration point been entered and
activated?
If not: Enter and activate calibration point(s). (
"calibration"
(*1)
function group)
• Check: Is the "reference pulse rate" (*1F) larger than 2
32
?
If yes: re-perform density calibration (
"calibration" (*1)
)
function group
A635
present date not defined
• Enter present date
(
"basic setup" (*0)
function group
"today's date" (*01)
function)
• See Note "
Error messages A165 "electronics defect" and
A635 "present date not defined"
on
ä
54.
A636
calibration date not plausible
Check calibration date and enter it again
(
"system parameters" (*C)
function group
"calibration date" (*C7)
function)
A637
operating mode not defined
Enter operating mode
("basic setup" (*0) function group
"operating mode" (*04)
function)
A638
measurement mode not defined
Enter measurement mode
("basic setup" (*0) function group
"measurement mode" (*05)
function)
A639
temperature compensation not
complete
Enter at least two "temperature - density" value pairs
(
"temperature compensation" (*3)
function group)
W662 high sensor temperature (warning)
Install water cooling jacket or thermal shielding
A663
sensor temperature too high (alarm)
Install water cooling jacket or thermal shielding
A664
temperature measurement error
Check correct functioning and connection of the PT-100 sensor
W681 current out of range
(3.8 to 20.5 mA)
Check calibration and linearization
A692
gammagraphy detected (alarm)
• Check, if interference radiation is present or the
"hold time"
(*54)
is too short.
• If no interference radiation is present: decrease
gammagraphy sensitivity
(
"gammagraphy" (*5)
function group
"sensitivity" (*52)
function)
W693 gammagraphy detected (warning)
Wait for end of the gammagraphy measurement
W695 measurement counter overflow
The local dose rate is too high (where applicable, reduce with
blind flange).
Code
Description
Remedy