5.6.5
Slot/index tables
Device Management
Analog-Input-Block
Parameter
Menu
position
Slot Index Size
[bytes]
Type
Read
Write Storage
Class
Dir. Object Header
1
0
12
STRING
X
constant
Comp. List Dir. Entry
1
1
24
STRING
X
constant
Parameter
Menu
position
Slot Index Size
[bytes]
Type
Read
Write Storage
Class
Standard parameters
Block Object
1
16
20
STRING
X
constant
Static revision
1
17
2
UINT16
X
static
Tag Description
1
18
32
STRING
X
X
static
Strategy
1
19
2
UINT16
X
X
static
Alert key
1
20
1
UINT8
X
X
static
Target Mode
1
21
1
ENUM8
X
X
static
Mode of Block
1
22
3
UINT8+
UINT8+
UINT8
X
dynamic
Alarm summary
1
23
8
UINT16
X
dynamic
Batch
1
24
10
UINT16
X
dynamic
Block parameters
Out
1
26
5
FLOAT+
UINT8
X
X
static
PV Scale
1
27
8
FLOAT+
FLOAT
X
X
static
Out Scale
1
28
11
FLOAT+
FLOAT+
UINT8
X
X
static
Linearization type
1
29
1
UINT8
X
X
static
Channel
1
30
2
UINT16
X
X
static
PV filter time
1
32
4
FLOAT
X
X
static
Fail safe type
1
33
1
UINT8
X
X
static
Fail safe value
1
34
4
FLOAT
X
X
static
Alarm Hysteresis
1
35
4
FLOAT
X
X
static
HI HI Limit
1
37
4
FLOAT
X
X
static
HI Limit
1
39
4
FLOAT
X
X
static
LO Limit
1
41
4
FLOAT
X
X
static
LO LO Limit
1
43
4
FLOAT
X
X
static