Commissioning
Gammapilot M FMG60 PROFIBUS PA
64
Hauser
6.4.2
Background calibration
Excerpt from the operating menu
The following excerpt from the operating menu shows how the background calibration is
entered. The individual functions are explained in the sections below.
"Background calibration" (*10)
Options:
•
stop/edit
This option must be selected if
– No background calibration is to be performed but the pulse rate of an existing
background calibration is to be displayed instead.
– A manual background calibration is to be performed.
After selecting this option, the Gammapilot M changes to the
"bgr. pulse rate" (*12)
function, where the existing pulse rate is displayed and can be changed if required.
•
start
This option is used to start an automatic background calibration. The Gammapilot M
changes to the
"avg. pulse rate" (*11)
function.
*10
Backgr. calibr.
0: stop/edit
1: start
*11
avg. pulse rate
*10
bgr. pulse rate
-1: not calibrated
On-site display
Meaning
Backgr. cal.
*10
This function is used to start the background calibration.
stop/edit
start
E
E
0