Commissioning
Gammapilot M FMG60 PROFIBUS PA
78
Hauser
6.5.4
Linearization (for concentration measurements)
If the concentration is to be measured in a unit other than the
"density unit" (*06)
, a
linearization must be performed after the calibration. This can be done in the
"Linearization" (*4)
function group. The functions of this group and the linearization
procedure are described in Operating Instructions BA00287F/00/EN, "Gammapilot M -
Description of Device Functions". This document can be found on the supplied CD-ROM.
6.5.5
Additional settings
After the calibration, the Gammapilot M outputs the measured value via the PROFIBUS PA
interface.
Many additional functions are available for optimizing the measuring point. They can be
configured as required. A detailed description of all instrument functions is given in
Operating Instructions BA00287F/00/EN, "Gammapilot M - Description of Device
Functions". This document can be found on the supplied CD-ROM.
6.6
Density measurement/temperature-compensated
Perform density measurement as described in the section "Calibration for density and
concentration measurements" and then perform temperature calibration (see "Temperature
compensation" section in BA00287F/00/EN "Gammapilot M - Description of Device
Functions").
6.7
Gammagraphy detection
See "Gammagraphy" section in BA00287F/00/EN "Gammapilot M - Description of Device
Functions".