6.1.1
Basic setup
Step
Function
Entry or selection
Comments
See page
1
Present date (*01)
e.g.
12.11.04 10:30
Set day with
O
and
S
;
Confirm with
F
;
The same applies for month, year,
hour and minute
ä
55
2
Beam type (*02)
• standard/cont.
• modulated
Select with
O
and
S
;
Confirm with
F
ä
55
3
Isotope (*03)
•
137
Cs
•
60
Co
• no compensation
Select with
O
and
S
;
Confirm with
F
ä
55
4
Operating mode (*04)
• Stand alone
• Master
• Slave
• End-slave
• not defined
Select with
O
and
S
;
Confirm with
F
ä
56
5
Measurement mode
(*05)
• level
• limit
• density
• concentration
Select with
O
and
S
;
Confirm with
F
ä
57
The following functions for density and concentration measurements only:
6
Density unit (*06)
• g/cm
3
• g/l
• ...
Select with
O
and
S
;
Confirm with
F
ä
58
7
Min. density (*07)
(lower limit of the
measuring range)
e.g.
0.9500 g/cm
3
Set each decimal place with
O
and
S
and confirm with
F
;
After the last place: confirm the
whole value with
F
.
ä
58
8
Max. density (*08)
(upper limit of the
measuring range)
e.g.
1.2500 g/cm
3
Set each decimal place with
O
and
S
and confirm with
F
;
After the last place: confirm the
whole value with
F
.
ä
58
9
Pipe diameter unit (*09) • mm
• inch
Select with
O
and
S
;
Confirm with
F
ä
58
10
Pipe diameter (*0A)
e.g.
200 mm
Set each decimal place with
O
and
S
and confirm with
F
;
After the last place: confirm the
whole value with
F
.
ä
59
The following function for all measurement methods:
11
Output damping (*0B)
e.g. 60 s
Set each decimal place with
O
and
S
and confirm with
F
;
After the last place: confirm the
whole value with
F
.
ä
60