Commissioning
Gammapilot M FMG60 PROFIBUS PA
52
Hauser
"Empty" or "free" calibration
For details see "Full and empty calibration or covered and free calibration",
ä
66.
Step
Function
Action/comments
1
• The radiation remains switched on
• Empty tank
Level: max. 40%
Level limit: beam path free
2
Calibration point level (*13)
Select "empty/free" with
O
or
S
;
confirm with
F
The following function for level measurement only
3
Value empty (*17)
Enter the current level (0% - 40%)
The following functions for level and level limit measurement
4
Calibration start stop (*15)
Select "start" with
O
or
S
;
confirm with
F
5
Avg. pulse rate (*11)
Wait until the value is stable, then confirm with
F
6
Calibration start stop (*15)
Select "stop/edit" with
O
or
S
;
confirm with
F
7
Empty calibration (*16)
Confirm value with
F
8
Calibration next point (*19)
Select "no" with
O
or
S
;
Confirm with
F
This completes the calibration.