Wiring
Gammapilot M FMG60 PROFIBUS PA
26
Hauser
4.9
Wiring in terminal compartment 2
CAUTION
!
Before connection please note the following:
‣
Connect potential matching line to transmitter ground terminal and to the ground
terminal of the water cooling jacket (if present) before connecting up the instrument
("Potential equalization",
ä
24).
‣
The cable isolations must comply with the supply voltage and the overvoltage category.
‣
The temperature resistance of the connecting cable must comply with the ambient
temperature.
The procedure
Unscrew the cover of the terminal
compartment.
Push the following cables through the
appropriate cable glands or threads:
- signal cable (if the signal output is
located in terminal compartment 2)
- PT-100 cable (if present)
- cascading cables (input and/or output,
if required)
Wire up according to the terminal
assignment diagram.
Tighten the cable glands or threads.
Screw the cover securely back onto the
terminal compartment.
A0018927
+ -
PT100
IN CASCADE OUT
SIM WP
OFF
ON
3
4
5 6
7 8
9 10 11 12
+ - - +
0
13
2