Troubleshooting
Gammapilot M FMG60 PROFIBUS PA
88
Hauser
9.1.4
Error codes
Code
Description
Remedy
A102
checksum error
Call Hauser Service
W103 initializing
Wait for completion of the initializing procedure
A106
downloading
Wait for completion of the download
A110
checksum error
Call Hauser Service
A111
electronics defect
Switch instrument off/on;
If error still is present: call Hauser service or exchange
transmitter
A113
electronics defect
Switch instrument off/on;
If error still is present: call Hauser service or exchange
transmitter
A114
electronics defect
Switch instrument off/on;
If error still is present: call Hauser service or exchange
transmitter
A116
download error
Repeat download
A121
electronics defect
Switch instrument off/on;
If error still is present: call Hauser service or exchange
transmitter
W153 initializing
Wait for completion of the initializing procedure
A160
checksum error
Call Hauser Service
A165
electronics defect
• Switch instrument off/on;
If error still is present: call Hauser service or
exchange transmitter
• See Note "
Error messages A165 "electronics defect" and
A635 "present date not defined"
on
ä
54.
A291
slave error
Check basic setup and connection of the slave transmitter
A503
wrong sensor type
Call Hauser Service
W513 calibration integration running
Wait, until a stable pulse rate has been reached; then,
terminate integration (by pressing "E" in the
"average pulse
rate" (*11)
function)
W514 PT-100 calibration
Wait for the calibration to complete.
if error still persists: call Hauser Service
A531
sensor electronics defect
Switch instrument off/on;
If error still is present: call Hauser service or exchange
transmitter
A532
sensor voltage error
Call Hauser Service
A533
wrong sensor software version
Call Hauser Service
A535
sensor regulation error
Call Hauser Service
W536 high voltage near limit
Call Hauser Service
A538
sensor communication error
Call Hauser Service
A602
linearization table not plausible
Check monotony of the linearization table; if necessary, adjust
table (
"linearization" (*4)
function group)
A612
linearization table not defined
Enter or complete linearization table
(
"linearization" (*4)
function group)
The linearization table must contain the end points 0% = 1000
cps (standardized) and 100% = 0 cps (standardized).
When entering via FieldCare:
Select the right type of table (lin. tab. "level" or "concentration")
W621 simulation on
Switch off linearization
(
"output" (*6)
function group
"simulation" (*65)
function)
W640 SIL lock device
SIL_locking not completed