Gammapilot M FMG60 PROFIBUS PA
Wiring
Hauser
25
4.8
Wiring in terminal compartment 1
CAUTION
!
Before connection please note the following:
‣
When using the instrument in hazardous areas, make sure to comply with national
standards and the specifications in the Safety Instructions (XAs). Make sure you use the
specified cable gland.
‣
The supply voltage must comply to the data on the nameplate.
‣
Switch off power supply before connecting the instrument.
‣
Connect potential matching line to the outer transmitter ground terminal and to the
ground terminal of the water cooling jacket (if present) before connecting up the
instrument (see "Potential equalization",
ä
24).
‣
Connect protective earth to the protective earth terminal (see "Terminal assignment",
ä
21).
‣
According to IEC/EN 61010 a suitable power switch has to be provided for the
instrument.
‣
The cable isolations must comply with the supply voltage and the overvoltage category.
‣
The temperature resistance of the connecting cable must comply with the ambient
temperature.
The procedure
Using a 3 mm Allen key, loosen the
cover clamp for the connection
compartment cover.
Unscrew the cover of the terminal
compartment.
Push the power cable and (if required)
the signal cable) through the
appropriate cable glands or cable
threads.
Wire up according to the terminal
assignment diagram.
Tighten the cable glands or threads.
Screw the cover securely back onto the
terminal compartment..
Adjust the cover clamp so it is set over
the cover and tighten.
A0019826
1
L1
L+
N
L-
+
-