Commissioning
Gammapilot M FMG60 PROFIBUS PA
76
Hauser
"Density calibration" (*1B)
"Density value" (*1C)
NOTICE
When entering the value, temperature influences have to be taken into account. The
density entered must refer to the temperature at which the pulse rate has been
determined. If the density and the pulse rate have been determined at different
temperatures, the density value must be corrected accordingly.
"Calibration point" (*1D)
Options:
•
not used
The calibration point is
not
used. However, it can be reactivated at a later point of time.
•
used
The calibration point is used.
•
clear
The calibration point is deleted. It cannot be reactivated at a later point of time.
On-site display
Meaning
density calibr.
*1B
The pulse rate of the respective calibration point is displayed in
this function. The display value must be confirmed by pressing
"E".
"-1"
indicates, that no pulse rate is present yet. In this case
there are two options:
• either return to the "calibration" (*15) function and restart
the calibration
• or enter a known or calculated pulse rate (manual calibra-
tion)
1983 cps
On-site display
Meaning
density value
*1C
This function is used to enter the density of the calibration
point. The value must be determined from the sample in a
laboratory measurement.
0.9963 g/cm3
On-site display
Meaning
calibr. point
*1D
This function is used to specify, if the current calibration piont
is to be used.
not used
used
clear