Gammapilot M FMG60 PROFIBUS PA
Operation
Hauser
33
5.2.2
The operating menu
Function code
The functions of the Gammapilot M are arranged in an operating menu. To ensure easy
orientation within the menu, a unique position code is indicated on the display for each
function. This code consists of one alphabetic and two numeric characters.
A0019876-EN
1
Measuring mode
2
Function group
3
Function
• The alphabetic character specifies the current measuring mode of the Gammapilot M:
–
L
: level
–
S
: limit (switch)
–
D
: density
–
C
: concentration
–
*
: no measuring mode selected yet
• The first numeric character identifies the function group:
–
basic setup *0
–
calibration *1
–
Safety settings *2
–
...
• The second numeric character numbers the individual functions within the function group:
basic setup *0
–
today's date *01
–
beam type *02
–
isotope *03
–
operating mode *04
–
...
Hereafter, the position is always given in brackets after the function name. "
*
" (not yet
selected) is always indicated as the measurement method, e.g.
"present date" (*01)
.