Gammapilot M FMG60 PROFIBUS PA
Commissioning
Hauser
55
6.3
Basic setup
6.3.1
"Present date" (*01)
6.3.2
"Beam type" (*02)
6.3.3
"Isotope" (*03)
On-site display
Meaning
Present date
*01
Date and time of the basic setup are specified in this function.
Each of these values must be confirmed by "E" after it has been
entered.
17.11.04
10:30
dd.mm.yy
hh:mm
On-site display
Meaning
Beam type
*02
This function is used to specify whether the radiation source
used emits radiation continuously or whether it is modulated
(for grammagraphy suppression).
• Standard/continuous (permanent, continuous radiation)
• Modulated (modulated radiation source)
Standard/cont.
Modulated
On-site display
Meaning
Isotope
*03
This function is used to specify which isotope is used for the
measurement. The Gammapilot M needs this information for
the decay compensation.
137 Cs
60 Co
no compens.