Maintenance instructions
Replacing the consumables in the O2 gas inlet
Fitting
1. Insert the filter (
Fig. 71
/1) in the O2 gas inlet.
2. Insert the sealing ring (
Fig. 70
/1) in the O2 gas inlet.
3. Use the screws (
Fig. 69
/1) to secure the connector plate (
Fig. 69
/2) to the
O2 gas inlet.
4. Insert the washer (
Fig. 68
/1) in the O2 gas inlet.
5. Push the seal (
Fig. 67
/1) onto the connector plate.
6. Screw the connection (
Fig. 66
/1) onto the O2 supply block with a tighten-
ing torque of 20 Nm +/-2 Nm using a torque spanner and insertion tool
(size 22 mm).
7. Fit the rear panel, see also „Disassembling/assembling the device“ sec-
tion headed „Rear panel“.
8. Re-attach removed tubing and connectors to the appropriate connections
on the Savina 300.
9. Fit the filter cover, see also „Disassembling/assembling the device“ sec-
tion headed „Filter cover“.
10. Perform the „electrical safety test“ and „function tests“ as per the test
instructions.
Fitting of the consumables is complete.
Содержание Savina 300
Страница 2: ...No 1309 0000001227_Publication 2 Savina 300 This page has been intentionally left blank ...
Страница 4: ...No 1309 0000001227_Publication 4 Savina 300 This page has been intentionally left blank ...
Страница 30: ...No 1309 0000001227_Publication 30 Savina 300 This page has been intentionally left blank ...
Страница 70: ...No 1309 0000001227_Publication 70 Savina 300 This page has been intentionally left blank ...
Страница 72: ...No 1309 0000001227_Publication 72 Savina 300 This page has been intentionally left blank ...
Страница 73: ...Parts catalog Savina 300 5664 920 Revision 01 2010 07 23 ...
Страница 110: ... This page has been intentionally left blank ...
Страница 112: ...No 0208_0000002533_Publication 2 50 6DYLQD 5HYLVLRQ This page has been intentionally left blank ...
Страница 164: ... This page has been intentionally left blank ...
Страница 165: ... This page has been intentionally left blank ...