3.2.1
Visual check
OK
3.2.2
Protective conductor resistance
Ohm
3.2.3
Earth leakage current
3.2.3.1
Normal condition (N.C.)
μA
3.2.3.2
Single fault condition (S.F.C.)
μA
3.2.3.3
Normal condition (N.C.)
μA
3.2.3.4
Single fault condition (S.F.C.)
μA
3.2.4
Patient leakage current
3.2.4.1
Normal condition (N.C.) DC
μA
3.2.4.2
Single fault condition (S.F.C.) DC
μA
3.2.4.3
Normal condition (N.C.) AC
μA
3.2.4.4
Single fault condition (S.F.C.) AC
μA
3.2.5
Optional multiple socket strip on trolley
3.2.5.1
Power fuse links
OK
3.2.5.2
Protective conductor resistance
Ohm
3.2.5.3
Equivalent device leakage current
μA
3.2.5.4
Recurrent test
μA
4 Function and condition test
4.1
Condition tests
4.1.1
Accompanying documents
OK
4.1.2
Device labelling
OK
4.1.3
General condition (Savina 300, accessories and
special accessories)
OK
4.1.4
O2 compressed gas connecting tube
OK
4.1.5
Option, trolley (general condition)
OK
4.1.6
Option, LPO (external oxygen supply)
4.1.6.1
O2 supply connector socket
OK
4.1.6.2
Protective cap 8415745 for connector socket
OK
4.2
Function tests
4.2.1
Non-return valve in expiratory valve
4.2.1.1
Expiratory valve 1
OK
4.2.1.2
Expiratory valve 2
OK
4.2.2
Safety valve (D1)
OK
4.2.3
Emergency respiratory valve (D2)
OK
4.2.4
Power supply unit
4.2.4.1
Electrical supply display
OK
4.2.4.2
Internal supply voltage (internal battery)
OK
4.2.4.3
Power failure alarm
OK
4.2.5
Emergency expiratory valve
OK
4.2.6
Testing the ventilation
4.2.6.1
O2 calibration
OK
4.2.6.2
VT=500 mL (expiratory valve 1)
mL
4.2.6.3
VT=500 mL (expiratory valve 2)
mL
4.2.7
O2 control (VC-CMV/VC-AC)
4.2.7.1
Measured value (external O2 meter)
vol% 02
4.2.8
Paw pressure regulation (VC-CMV/VC-AC)
4.2.8.1
Expiratory phase, 5 mbar (external pressure
gauge)
mbar
4.2.8.2
Expiratory phase, 25 mbar (external pressure
gauge)
mbar
4.3
Options
4.3.1
AWT 01 temperature sensor
OK
4.3.2
Nurse call
OK
4.3.3
External battery or DC onboard power
4.3.3.1
Function tested with external battery
OK
4.3.3.2
Function tested with DC onboard power
OK
4.3.4
LPO (external oxygen supply)
OK
4.4
User-specific settings
4.4.1
Input of user-specific settings
OK
OK
Result
4.5
Final action
4.5.1
Device handover with test label
OK
5 Test equipment
5.1
Test equipment list
5.1.1
Test equipment subject to mandatory calibration
5.1.1.1
Test equipment calibrated in a valid manner
used.
OK
OK
Result
Содержание Savina 300
Страница 2: ...No 1309 0000001227_Publication 2 Savina 300 This page has been intentionally left blank ...
Страница 4: ...No 1309 0000001227_Publication 4 Savina 300 This page has been intentionally left blank ...
Страница 30: ...No 1309 0000001227_Publication 30 Savina 300 This page has been intentionally left blank ...
Страница 70: ...No 1309 0000001227_Publication 70 Savina 300 This page has been intentionally left blank ...
Страница 72: ...No 1309 0000001227_Publication 72 Savina 300 This page has been intentionally left blank ...
Страница 73: ...Parts catalog Savina 300 5664 920 Revision 01 2010 07 23 ...
Страница 110: ... This page has been intentionally left blank ...
Страница 112: ...No 0208_0000002533_Publication 2 50 6DYLQD 5HYLVLRQ This page has been intentionally left blank ...
Страница 164: ... This page has been intentionally left blank ...
Страница 165: ... This page has been intentionally left blank ...