N
o
.02
08
_00
00
00
253
3_
Pu
bl
ic
ati
on
6DYLQD
Electrical safety
Electrical safety according to IEC 60601-1
26 / 50
6DYLQD
5HYLVLRQ
3.2.3
Earth leakage current
Test set-up
Fig. 6
Earth leakage current test setup
Action
• Prepare the test setup.
• Switch on the tester and the device under test.
• Follow the instructions on the tester.
Test
Normal condition (N.C.): The value must not exceed
500
μA.
Entry
Normal condition (N.C.)
[________μA]
Test
Single fault condition (S.F.C.): The value must not exceed
1000
μA.
Entry
Single fault condition (S.F.C.)
[________μA]
Action
• Plug the power supply connector (inverted, if possible) into the test socket
of the test device. (In many test devices the power supply connector inver-
sion can be simulated by means of a built-in selector switch.)
Test
Normal condition (N.C.): The value must not exceed
500
μA.
NOTE
In order to avoid incorrect measurement, set up the device under test so that
it is insulated.
Item Designation
1
Tester, e.g. SECUTEST
2
Device under test
L
Conductor
N
Neutral conductor
PE
(Protective earth) conductor
NOTE
For symmetrical mains plugs that have no preferential position in the socket-
outlet, the earth leakage current test must be performed twice! The second
test is performed with the plug rotated 180° in the socket. In many test de-
vices the mains plug rotation is simulated by means of a built-in selector
switch. The higher measured value must be documented.
Содержание Savina 300
Страница 2: ...No 1309 0000001227_Publication 2 Savina 300 This page has been intentionally left blank ...
Страница 4: ...No 1309 0000001227_Publication 4 Savina 300 This page has been intentionally left blank ...
Страница 30: ...No 1309 0000001227_Publication 30 Savina 300 This page has been intentionally left blank ...
Страница 70: ...No 1309 0000001227_Publication 70 Savina 300 This page has been intentionally left blank ...
Страница 72: ...No 1309 0000001227_Publication 72 Savina 300 This page has been intentionally left blank ...
Страница 73: ...Parts catalog Savina 300 5664 920 Revision 01 2010 07 23 ...
Страница 110: ... This page has been intentionally left blank ...
Страница 112: ...No 0208_0000002533_Publication 2 50 6DYLQD 5HYLVLRQ This page has been intentionally left blank ...
Страница 164: ... This page has been intentionally left blank ...
Страница 165: ... This page has been intentionally left blank ...