Function descriptions
Pneumatic assembly
Flow sensor S1 measures the inspiratory gas flow. The measurement vari-
able is used to calculate the necessary oxygen flow and to actuate the oxy-
gen metering valves V7.1 to V7.8 in order to control the breaths and monitor
the device functions. The flow sensor includes a temperature measurement
function to measure the inspiratory gas temperature.
Fig. 19
Detail view of functional diagram; flow sensor
The flow sensor S2 measures the gas flow through the expiratory valve. The
flow sensor is a temperature-compensated hot-wire anemometer with no flow
direction detector. With this signal the patient is monitored (e.g. the minute
volume).
Fig. 20
Detail view of functional diagram; flow sensor
Expiration is executed with the directly actuated expiratory valve V3.
The expiratory valve has the following functions:
–
PEEP control during the expiratory phase
–
Close the breathing system during the inspiratory phase.
Содержание Savina 300
Страница 2: ...No 1309 0000001227_Publication 2 Savina 300 This page has been intentionally left blank ...
Страница 4: ...No 1309 0000001227_Publication 4 Savina 300 This page has been intentionally left blank ...
Страница 30: ...No 1309 0000001227_Publication 30 Savina 300 This page has been intentionally left blank ...
Страница 70: ...No 1309 0000001227_Publication 70 Savina 300 This page has been intentionally left blank ...
Страница 72: ...No 1309 0000001227_Publication 72 Savina 300 This page has been intentionally left blank ...
Страница 73: ...Parts catalog Savina 300 5664 920 Revision 01 2010 07 23 ...
Страница 110: ... This page has been intentionally left blank ...
Страница 112: ...No 0208_0000002533_Publication 2 50 6DYLQD 5HYLVLRQ This page has been intentionally left blank ...
Страница 164: ... This page has been intentionally left blank ...
Страница 165: ... This page has been intentionally left blank ...