Maintenance instructions
Replacing the O2 sensors
Calibrating the O2 sensors
Preconditions
–
The Savina 300 has been connected to the mains power supply.
–
The blade fuse has been has been inserted into the slot on the power
supply unit.
–
The Savina 300 has been connected to the O2 supply.
Calibrating O2 sensor 2 in
„HPO“ mode
1. Switch on the Savina 300.
2. Set „VC-CMV/VC-AC“ mode.
3. Press the „Sensors/Parameters“ softkey.
4. Press the O2 calibration „Start“ softkey and confirm with the rotary knob.
The message „Disconnect patient“ is displayed.
5. Within the next 30 seconds detach the inspiratory tube from the inspira-
tory socket.
The message „O2 calibration in progress“ is displayed.
6. Detach the inspiratory tube from the inspiratory socket and, if necessary,
continue ventilation with an independent ventilator.
The Savina 300 calibrates O2 sensor 2. During calibration the alarms which
would normally occur due to the disconnection and the changed O2 concen-
tration are disabled. After about 60 seconds the prompt „Reconnect patient“
is displayed.
7. Immediately fit the inspiratory tube on the inspiratory socket.
If the display indicates „O2 calibration OK“, calibration is complete.
NOTE
To be able to calibrate the Savina 300 correctly, you must wait for the
device to warm up. If the blade fuse was removed, it may take as long as
20 minutes for the device to warm up.
CAUTION
If the quality of the oxygen from the central oxygen supply or the com-
pressed gas oxygen cylinder is inadequate, calibrate the O2 sensor with
calibration gas (100% O2). Otherwise miscalibration may occur.
NOTE
If the patient has not been reconnected after 30 seconds, Savina 300 starts
ventilating again in the preset ventilation mode and all alarms are enabled
again.
NOTE
If the message „FiO2 measurement failed“ is displayed after calibrating, the
O2 sensors must be replaced – see „Replacing the O2 sensors“.
Содержание Savina 300
Страница 2: ...No 1309 0000001227_Publication 2 Savina 300 This page has been intentionally left blank ...
Страница 4: ...No 1309 0000001227_Publication 4 Savina 300 This page has been intentionally left blank ...
Страница 30: ...No 1309 0000001227_Publication 30 Savina 300 This page has been intentionally left blank ...
Страница 70: ...No 1309 0000001227_Publication 70 Savina 300 This page has been intentionally left blank ...
Страница 72: ...No 1309 0000001227_Publication 72 Savina 300 This page has been intentionally left blank ...
Страница 73: ...Parts catalog Savina 300 5664 920 Revision 01 2010 07 23 ...
Страница 110: ... This page has been intentionally left blank ...
Страница 112: ...No 0208_0000002533_Publication 2 50 6DYLQD 5HYLVLRQ This page has been intentionally left blank ...
Страница 164: ... This page has been intentionally left blank ...
Страница 165: ... This page has been intentionally left blank ...