Function and condition test
Function tests
38 / 50
6DYLQD
5HYLVLRQ
4.2.6
Testing the ventilation
Prerequisites
The Savina 300 is connected to the mains power supply and to the „O2“ com-
pressed gas supply and switched on.
Action
• Prepare the following test setup:
Test set-up
Fig. 12
Test setup; testing the ventilation
Action
• Press the „Ventilation settings“ softkey and select „VC-CMV/VC-AC” mode.
• Set the following parameters:
– O2: 21%
– VT: 500 mL
– Ti : 2.0 s
– f: 12/min
– PEEP: 5 mbar
• Press the „Advanced settings“ softkey and set the following parameters:
– Sigh: Off
– Flow trigger: Off
– AutoFlow: Off
Item
Designation
1
Pressure gauge
2
Test lung
3
Y-piece, adult, Luer-Lock
4
Adult ventilation tubing
5
O2 analyzer
6
O2 adapter
Содержание Savina 300
Страница 2: ...No 1309 0000001227_Publication 2 Savina 300 This page has been intentionally left blank ...
Страница 4: ...No 1309 0000001227_Publication 4 Savina 300 This page has been intentionally left blank ...
Страница 30: ...No 1309 0000001227_Publication 30 Savina 300 This page has been intentionally left blank ...
Страница 70: ...No 1309 0000001227_Publication 70 Savina 300 This page has been intentionally left blank ...
Страница 72: ...No 1309 0000001227_Publication 72 Savina 300 This page has been intentionally left blank ...
Страница 73: ...Parts catalog Savina 300 5664 920 Revision 01 2010 07 23 ...
Страница 110: ... This page has been intentionally left blank ...
Страница 112: ...No 0208_0000002533_Publication 2 50 6DYLQD 5HYLVLRQ This page has been intentionally left blank ...
Страница 164: ... This page has been intentionally left blank ...
Страница 165: ... This page has been intentionally left blank ...