N
o
.13
09
_00
00
00
122
7_
Pu
bl
ic
ati
on
42
Savina 300
Maintenance instructions
Replacing the O2 sensors
Replacing the O2 sensors
Introduction
The following section provides a view of the O2 sensors and describes how
to remove, fit and calibrate them.
View
The O2 sensors (
Fig. 35
/1) are located behind the cover plate on the front of
the Savina 300.
Fig. 35
Device front; O2 sensors
NOTE
Replace the O2 sensors if calibration is no longer possible or if the alarm
message „FiO2 measurement failed“ is displayed.