Electrical safety
Electrical safety to EN 62353
16 / 50
6DYLQD
5HYLVLRQ
3.1
Electrical safety to EN 62353
Introduction
The following subsections provide descriptions of device checks, recurrent
testing and testing after servicing of medical electrical (ME) devices.
3.1.1
Visual check
Prerequisites
The tester and the device under test are switched off.
Action
• Disconnect the power plug from the mains socket.
Test
– The power fuse-links of the device under test match the specifications on
the rating plate.
– The power cable and plug are not dirty or damaged.
Result
Condition checked.
[________OK]
NOTE
The device conforms to the conditions of protection class I. When the airway
temperature sensor is connected it conforms to type BF.
NOTE
The tester, e.g. SECUTEST, must be correctly configured for all measure-
ments. If implausible measurement results are obtained, such as a leakage
current of 0.0 μA, check the tester configuration in addition to the test setup!
NOTE
– In testing to IEC 62353, the medical electrical device (ME device) or the
medical electrical system (ME system) must be tested.
– ME systems must be treated like ME devices.
– An ME system is a combination of several devices, as specified by the
manufacturer, of which at least one must be an ME device, which are
interconnected by a functional connection or by means of a multiple
socket outlet.
NOTE
In the case of devices connected to other devices by a data cable, this con-
nection must be disconnected prior to performing the electrical safety test,
in order to avoid false measurements.
WARNING!
Hazardous voltage.
Touching live components can lead to serious injury or death.
– Disconnect the power cord from the AC outlet before checking the power
fuse-links.
Содержание Savina 300
Страница 2: ...No 1309 0000001227_Publication 2 Savina 300 This page has been intentionally left blank ...
Страница 4: ...No 1309 0000001227_Publication 4 Savina 300 This page has been intentionally left blank ...
Страница 30: ...No 1309 0000001227_Publication 30 Savina 300 This page has been intentionally left blank ...
Страница 70: ...No 1309 0000001227_Publication 70 Savina 300 This page has been intentionally left blank ...
Страница 72: ...No 1309 0000001227_Publication 72 Savina 300 This page has been intentionally left blank ...
Страница 73: ...Parts catalog Savina 300 5664 920 Revision 01 2010 07 23 ...
Страница 110: ... This page has been intentionally left blank ...
Страница 112: ...No 0208_0000002533_Publication 2 50 6DYLQD 5HYLVLRQ This page has been intentionally left blank ...
Страница 164: ... This page has been intentionally left blank ...
Страница 165: ... This page has been intentionally left blank ...