6DYLQD
5HYLVLRQ
No
.0
208
_0
00
000
25
33
_Pu
bl
ic
ati
o
n
39 / 50
Function and condition test
Function tests
Action
• Press the „Alarms“ softkey and set the following parameters:
– MV high: 41 L/min
– MV low: 0.5 L/min
– Paw: 100 mbar
– VTi: 4.00 L
– f: 120/min
– Tapn: 60 s
Action
• Press the „Sensors/Parameters“ softkey and set the following parameters:
– O2 monitoring: On
– Flow monitoring: On
Action
• Press the „System configuration“ softkey.
• Press the „Ventilation“ softkey and set the following parameters:
– Pressure limitation: Off
– Plateau: On
– LPO: Off
• Start ventilation with the Savina 300.
O2 calibration
Action
• Select the „Sensoren/Parameter” menu.
• Press the O2 calibration „Start“ softkey and confirm with the rotary knob.
Test
The message „Disconnect patient“ is displayed.
Action
• Disconnect the ventilation tube from the inspiratory socket.
Test
The display shows „O2 calibration in progress“ and then „Reconnect patient“.
Action
• Reconnect the ventilation tube to the inspiratory socket.
Test
The following message is displayed: „O2 calibration OK“
Entry
[________OK]
VT flow control and VT measurement
Action
• While the test lung is being ventilated, push the Spirolog sensor to the left-
hand end position and back to the right-hand end position in order to trig-
ger a flow calibration.
• After the display message „Flow calibration OK“ appears, ventilate the test
lung for at least another three cycles.
• Read out the „VTe“ measured value and compare it with the specified toler-
ance range.
VT=500 mL (expiratory valve 1)
The displayed value is in the range from
440
mL to
540
mL.
Entry
[________mL]
Action
• After the tidal volume measurement (VT) perform a flow calibration.
NOTE
Perform the „VT metering and VT measurement“ test with
all
the expiratory
valves listed in the „Device configuration“ section under „Serial numbers“.
Содержание Savina 300
Страница 2: ...No 1309 0000001227_Publication 2 Savina 300 This page has been intentionally left blank ...
Страница 4: ...No 1309 0000001227_Publication 4 Savina 300 This page has been intentionally left blank ...
Страница 30: ...No 1309 0000001227_Publication 30 Savina 300 This page has been intentionally left blank ...
Страница 70: ...No 1309 0000001227_Publication 70 Savina 300 This page has been intentionally left blank ...
Страница 72: ...No 1309 0000001227_Publication 72 Savina 300 This page has been intentionally left blank ...
Страница 73: ...Parts catalog Savina 300 5664 920 Revision 01 2010 07 23 ...
Страница 110: ... This page has been intentionally left blank ...
Страница 112: ...No 0208_0000002533_Publication 2 50 6DYLQD 5HYLVLRQ This page has been intentionally left blank ...
Страница 164: ... This page has been intentionally left blank ...
Страница 165: ... This page has been intentionally left blank ...