Function descriptions
Pneumatic assembly
O2 mixture with O2 high
pressure
In order to be able to ventilate with an increased O2 concentration, the unit
must be supplied with 2.7 to 6.0 bar O2. The oxygen is filtered by the filter F2.
With the aid of the digital valve block consisting of 8 digital solenoid valves,
oxygen is metered into the volume (mixing chamber). The amount of metered
oxygen depends on the pre-set O2 concentration and on the inspiratory flow
rate measured by the flow sensor S1. The oxygen is metered in a closed con-
trol loop. In the process, the inspiratory O2 concentration is measured by the
O2 sensor S 3.1.
Fig. 11
Detail view of functional diagram; O2 mixture with O2 high pres-
sure
Содержание Savina 300
Страница 2: ...No 1309 0000001227_Publication 2 Savina 300 This page has been intentionally left blank ...
Страница 4: ...No 1309 0000001227_Publication 4 Savina 300 This page has been intentionally left blank ...
Страница 30: ...No 1309 0000001227_Publication 30 Savina 300 This page has been intentionally left blank ...
Страница 70: ...No 1309 0000001227_Publication 70 Savina 300 This page has been intentionally left blank ...
Страница 72: ...No 1309 0000001227_Publication 72 Savina 300 This page has been intentionally left blank ...
Страница 73: ...Parts catalog Savina 300 5664 920 Revision 01 2010 07 23 ...
Страница 110: ... This page has been intentionally left blank ...
Страница 112: ...No 0208_0000002533_Publication 2 50 6DYLQD 5HYLVLRQ This page has been intentionally left blank ...
Страница 164: ... This page has been intentionally left blank ...
Страница 165: ... This page has been intentionally left blank ...