Function and condition test
Function tests
4.2.3
Emergency respiratory valve (D2)
Prerequisites
The Savina 300 is not connected to the mains power supply or to the „O2“
compressed gas supply. Savina 300 is switched off.
Action
• Prepare the following test setup:
Test set-up
Fig. 10
„Emergency respiratory valve“ test setup
Action
• Using the test pressure regulator, set a flow of 57 to 63 L/min.
Test
The measured pressure is in a range from
-3
mbar to
-6
mbar.
Entry
[________OK]
Item
Designation
1
Test pressure regulator
2
Injector
3
Silicone tube
4
Connecting sleeve
5
ISO socket
6
Flowmeter
7
Adult ventilation tube (tube length 0.9 to 1.2 m)
8
Sealing plug
9
Y-piece, adult, Luer-Lock
10
Silicone tube
11
Pressure gauge
Содержание Savina 300
Страница 2: ...No 1309 0000001227_Publication 2 Savina 300 This page has been intentionally left blank ...
Страница 4: ...No 1309 0000001227_Publication 4 Savina 300 This page has been intentionally left blank ...
Страница 30: ...No 1309 0000001227_Publication 30 Savina 300 This page has been intentionally left blank ...
Страница 70: ...No 1309 0000001227_Publication 70 Savina 300 This page has been intentionally left blank ...
Страница 72: ...No 1309 0000001227_Publication 72 Savina 300 This page has been intentionally left blank ...
Страница 73: ...Parts catalog Savina 300 5664 920 Revision 01 2010 07 23 ...
Страница 110: ... This page has been intentionally left blank ...
Страница 112: ...No 0208_0000002533_Publication 2 50 6DYLQD 5HYLVLRQ This page has been intentionally left blank ...
Страница 164: ... This page has been intentionally left blank ...
Страница 165: ... This page has been intentionally left blank ...