n
35
Maintenance instructions
Disassembling/assembling the device
Rear panel
Preconditions
–
Savina 300 has been disconnected from the O2 supply and the mains
power supply.
–
The tubing system and units mounted on the Savina 300 have been
removed.
–
EGB conditions to protect against electrostatic discharge have been
established.
–
The filter cover has been removed, see section headed „Filter cover“.
Removal
1. Pry the tab fuse (
Fig. 25
/2) from its holder using a small screwdriver or
similar tool.
2. Remove the connectors and tubing listed below from the Savina 300, if
fitted:
Fig. 25
View of device rear; connections
Item
Designation
1
Connection (external battery)
2
Tab fuse
3
Mains connection
4
Connection (nurse call)
5
Connection (RS232 port)
6
Connection for LPO (option)
7
Connection (O2 gas supply)
Connector for potential equalization
Содержание Savina 300
Страница 2: ...No 1309 0000001227_Publication 2 Savina 300 This page has been intentionally left blank ...
Страница 4: ...No 1309 0000001227_Publication 4 Savina 300 This page has been intentionally left blank ...
Страница 30: ...No 1309 0000001227_Publication 30 Savina 300 This page has been intentionally left blank ...
Страница 70: ...No 1309 0000001227_Publication 70 Savina 300 This page has been intentionally left blank ...
Страница 72: ...No 1309 0000001227_Publication 72 Savina 300 This page has been intentionally left blank ...
Страница 73: ...Parts catalog Savina 300 5664 920 Revision 01 2010 07 23 ...
Страница 110: ... This page has been intentionally left blank ...
Страница 112: ...No 0208_0000002533_Publication 2 50 6DYLQD 5HYLVLRQ This page has been intentionally left blank ...
Страница 164: ... This page has been intentionally left blank ...
Страница 165: ... This page has been intentionally left blank ...