
N
o
.13
09
_00
00
00
122
7_
Pu
bl
ic
ati
on
64
Savina 300
Maintenance instructions
Replacing the external batteries
14. Secure the cover by the screw (
Fig. 64
/1).
Fig. 64
Detail view of trolley; screw
15. Insert the connection socket of the DS battery cable for the external bat-
teries in the connector on the rear of the Savina 300.
Fitting of the external batteries is complete.
16. Perform the „function test“ and „electrical safety test“ as per the test
instructions.
17. Fully charge the external batteries (for time taken see instructions for
use).
CAUTION
Batteries not fully charged! If the batteries are not fully charged, the running
time of the device in the event of a mains power failure may be reduced
considerably.
–
Inform the user that the rechargeable batteries still need to be
recharged.
Содержание Savina 300
Страница 2: ...No 1309 0000001227_Publication 2 Savina 300 This page has been intentionally left blank ...
Страница 4: ...No 1309 0000001227_Publication 4 Savina 300 This page has been intentionally left blank ...
Страница 30: ...No 1309 0000001227_Publication 30 Savina 300 This page has been intentionally left blank ...
Страница 70: ...No 1309 0000001227_Publication 70 Savina 300 This page has been intentionally left blank ...
Страница 72: ...No 1309 0000001227_Publication 72 Savina 300 This page has been intentionally left blank ...
Страница 73: ...Parts catalog Savina 300 5664 920 Revision 01 2010 07 23 ...
Страница 110: ... This page has been intentionally left blank ...
Страница 112: ...No 0208_0000002533_Publication 2 50 6DYLQD 5HYLVLRQ This page has been intentionally left blank ...
Страница 164: ... This page has been intentionally left blank ...
Страница 165: ... This page has been intentionally left blank ...