Function and condition test
Function tests
Entry
[________OK]
4.2.5
Emergency expiratory valve
Prerequisites
The Savina 300 is connected to the mains power supply and to the „O2“ com-
pressed gas supply. Savina 300 is switched on.
Action
• Press the „Ventilation settings“ softkey and select „VC-CMV/VC-AC” mode.
• Set the following parameters:
– O2: 21%
– VT: 500 mL
– Ti : 5.0 s
– f: 6/min
– PEEP: 5 mbar
Action
• Press the „Alarms“ softkey
– Airway pressure limit high (Paw high): Set 10 mbar above peak pressure
Action
• Prepare the following test setup:
Fig. 11
Test setup; „emergency expiratory valve“
Test
After three cycles the Savina 300 builds up a pressure up to the airway pres-
sure limit (Paw high). Then the pressure is briefly relieved by emergency
expiratory valve to less than 6 mbar.
Entry
[________OK]
Item
Designation
1
Pressure gauge
2
Silicone tube
3
Connecting sleeve
Содержание Savina 300
Страница 2: ...No 1309 0000001227_Publication 2 Savina 300 This page has been intentionally left blank ...
Страница 4: ...No 1309 0000001227_Publication 4 Savina 300 This page has been intentionally left blank ...
Страница 30: ...No 1309 0000001227_Publication 30 Savina 300 This page has been intentionally left blank ...
Страница 70: ...No 1309 0000001227_Publication 70 Savina 300 This page has been intentionally left blank ...
Страница 72: ...No 1309 0000001227_Publication 72 Savina 300 This page has been intentionally left blank ...
Страница 73: ...Parts catalog Savina 300 5664 920 Revision 01 2010 07 23 ...
Страница 110: ... This page has been intentionally left blank ...
Страница 112: ...No 0208_0000002533_Publication 2 50 6DYLQD 5HYLVLRQ This page has been intentionally left blank ...
Страница 164: ... This page has been intentionally left blank ...
Страница 165: ... This page has been intentionally left blank ...