Function and condition test
Function tests
34 / 50
6DYLQD
5HYLVLRQ
4.2.2
Safety valve (D1)
Prerequisites
The Savina 300 is not connected to the mains power supply or to the „O2“
compressed gas supply. Savina 300 is switched off.
Action
• Prepare the following test setup:
Test set-up
Fig. 9
Test setup; „pneumatic safety valve“
Action
• Using a test pressure regulator, set a flow of 2 to 3 L/min.
Test
The measured pressure is in a range from
80
mbar to
120
mbar.
Entry
[________OK]
Item
Designation
1
Test pressure regulator
2
Flowmeter
3
Pressure gauge
4
T-piece
5
Silicone tube
6
Connecting sleeve
Содержание Savina 300
Страница 2: ...No 1309 0000001227_Publication 2 Savina 300 This page has been intentionally left blank ...
Страница 4: ...No 1309 0000001227_Publication 4 Savina 300 This page has been intentionally left blank ...
Страница 30: ...No 1309 0000001227_Publication 30 Savina 300 This page has been intentionally left blank ...
Страница 70: ...No 1309 0000001227_Publication 70 Savina 300 This page has been intentionally left blank ...
Страница 72: ...No 1309 0000001227_Publication 72 Savina 300 This page has been intentionally left blank ...
Страница 73: ...Parts catalog Savina 300 5664 920 Revision 01 2010 07 23 ...
Страница 110: ... This page has been intentionally left blank ...
Страница 112: ...No 0208_0000002533_Publication 2 50 6DYLQD 5HYLVLRQ This page has been intentionally left blank ...
Страница 164: ... This page has been intentionally left blank ...
Страница 165: ... This page has been intentionally left blank ...