Savina 300
No
.1
309
_0
00
000
12
27
_Pu
bl
ic
ati
o
n
43
Maintenance instructions
Replacing the O2 sensors
Removal
Preconditions
–
The Savina 300 is switched off and disconnected from the mains power
supply.
Service equipment
required
–
Screwdriver
Procedure
1. Swivel the inspiratory port (
Fig. 36
/1) down.
2. Loosen the screw (
Fig. 36
/2) using a screwdriver or similar and detach
the cover plate (
Fig. 36
/3).
Fig. 36
Device front; cover plate
3. Take the O2 sensors (
Fig. 37
/1) out of their mounts.
Fig. 37
Device front; removing the O2 sensors
NOTE
O2 sensors are special waste. When disposing of them, observe local
waste disposal regulations.
Содержание Savina 300
Страница 2: ...No 1309 0000001227_Publication 2 Savina 300 This page has been intentionally left blank ...
Страница 4: ...No 1309 0000001227_Publication 4 Savina 300 This page has been intentionally left blank ...
Страница 30: ...No 1309 0000001227_Publication 30 Savina 300 This page has been intentionally left blank ...
Страница 70: ...No 1309 0000001227_Publication 70 Savina 300 This page has been intentionally left blank ...
Страница 72: ...No 1309 0000001227_Publication 72 Savina 300 This page has been intentionally left blank ...
Страница 73: ...Parts catalog Savina 300 5664 920 Revision 01 2010 07 23 ...
Страница 110: ... This page has been intentionally left blank ...
Страница 112: ...No 0208_0000002533_Publication 2 50 6DYLQD 5HYLVLRQ This page has been intentionally left blank ...
Страница 164: ... This page has been intentionally left blank ...
Страница 165: ... This page has been intentionally left blank ...