Troubleshooting
5
Millenia / Indico 100 Series Service Manual Ch # 740904-09 Rev. B
Page 5-5
5.2.3 Error
Messages
(Cont)
ERROR CODE
MESSAGE
PROBLEM
ACTION
E016
TOMO BUT ERROR
Tomo exposure exceeded
back up time.
1. Check exposure
technique
settings.
2. Increase tomo
back up time if
necessary.
E017
NOT CALIBRATED
Selected mA not calibrated
for selected kV.
Recalibrate X-ray
tube.
E018
PREP TIMEOUT
Generator has been in prep
state too long.
Reduce length of
time in prep state.
E019 ANODE
HEAT
LIMIT
Selected
parameters
will
cause X-ray tube to exceed
its programmed anode heat
limit.
Reduce parameters
or wait for tube to
cool.
E020
THERMAL INT #1
X-ray tube # 1 too hot and
its thermal switch has
opened.
Wait for X-ray tube #
1 to cool.
E021
THERMAL INT #2
X-ray tube # 2 too hot and
its thermal switch has
opened.
Wait for X-ray tube #
2 to cool.
E022 DOOR
INTERLOCK
Door
is
open.
Close
door.
E023
COLLIMATOR ERR
Collimator is not ready.
Check collimator.
E024
CASSETTE ERROR
Cassette is not ready.
Check cassette.
E025
II SAFETY INT
II safety is not ready.
Check II safety.
E026
SPARE INT
Spare input is not ready.
Check spare input.
E028
PREP SW CLOSED
Prep input active during
power on initialization
phase.
Check prep switch
and input for short
circuit.
E029
X-RAY SW CLOSED
X-ray input active during
power on initialization
phase.
Check X-ray switch
and input for short
circuit.
E030
FLUORO SW CLOSED
Fluoro input active during
power on initialization
phase.
Check fluoro switch
and input for short
circuit.
E031 REMOTE
COMM
ERR
Communication
error
detected with remote fluoro
control unit.
1. Check remote
fluoro control cable
for damage and
proper connection.
2. Turn power off
and then on to
Reset Generator.