Acceptance Testing
4
Millenia / Indico 100 Series Service Manual Ch # 740904-08 Rev. F
Page 4-5
4.4.1
Console Rad Tests (Cont)
Step
Action
Result
Check
14.
Ensure that AEC is selected.
Press the FIELD button (23 X 56 cm console)
or press the individual AEC field select buttons
in sequence (31 X 42 cm console).
The three field indicator LED’s
light to indicate field selection
[L+C+R], [R], [C], [R+C], [L],
[L+R], [L+C] as the switch is
toggled (23 X 56 cm console).
For the 31 X 42 cm console, the
left, center, right field selection
LED’s should light as each field is
selected.
15.
Press power OFF button on console.
Unit switches off.
4.4.2 Console
Fluoro
Tests
This section applies only to R&F units.
BEFORE CONTINUING, ENSURE THAT THE REMOTE FLUORO CONTROL IS CONNECTED (IF
USED WITH THIS INSTALLATION).
Step
Action
Result
Check
1.
Press power ON button on console.
Unit switches on.
2.
Select an active fluoro receptor 1 to 6.
A: Fluoro display area of console
lights.
B: Remote fluoro control panel
lights (if used).
3.
Press the DOSE button 17.
Dose display on LCD display
changes as switch is toggled.
4.
Press the MAG button 20 on the console.
Press the MAG +/
-
buttons on the remote
fluoro control if used.
A: Mag display on console LCD
and remote fluoro display
changes (IF I/I MODES
ENABLED DURING
GENERATOR
CONFIGURATION) as switch
20 is toggled.
B: Mag display on console LCD
and remote fluoro display
increases if MAG + button is
pressed, decreases if MAG
-
button on the remote fluoro
control is pressed (IF I/I
MODES ENABLED DURING
GENERATOR
CONFIGURATION.
5.
Press the ABS button 19 on the Console to
enter ABS mode.
A: The LED adjacent to button
19 lights.
B: The LED adjacent to the ABS
button on the remote fluoro
control if used lights.
6.
Press the ABS button on the remote fluoro
control if used, else press ABS button 19 on
the console.
ABS indicator LED’s adjacent to
ABS buttons on console and
remote fluoro control if used are
off.