Reproducibility, Linearity, & HVL Testing
CPI Canada Inc
Page 8 Rev. A X-Ray Generator Service Manual Supplement 740917-00
4.0 LINEARITY
(Cont)
Record the mAs and mR values taken from tables 3, 4, 6, and 7 below.
Table 3
mAs
3
= _____
mR
3
= _____
Table 4
mAs
4
= _____
mR
4
= _____
Table 6
mAs
6
= _____
mR
6
= _____
Table 7
mAs
7
= _____
mR
7
= _____
mR
3
X
3
= = ________
mAs
3
mR
6
X
6
= = ________
mAs
6
X
3
- X
6
L = = ________ (not to exceed 0.095)
X
3
+ X
6
In the numerator of the above equation, use the absolute value of X
3
- X
6
(disregard the minus sign).
Table
8:
Linearity
mR
4
X
4
= = ________
mAs
4
mR
7
X
7
= = ________
mAs
7
X
4
- X
7
L = = ________ (not to exceed 0.095)
X
4
+ X
7
In the numerator of the above equation, use the absolute value of X
4
- X
7
(disregard the minus sign).
Table
9:
Linearity