4
Acceptance Testing
CPI Canada Inc.
Page 4-6
Rev. F Millenia / Indico 100 Series Service Manual Ch # 740904-08
4.4.2
Console Fluoro Tests (Cont)
Step
Action
Result
Check
7.
Press the fluoro kV +/
-
buttons on the console
and remote fluoro control if used.
KV increases if kv + pressed.
KV decreases if kv
-
pressed.
Confirm tracking of kV displays
on console and remote fluoro
control.
8.
Press the fluoro mA +/
-
buttons on the console
and remote fluoro control if used.
MA increases if mA + pressed.
MA decreases if mA
-
pressed.
Confirm tracking of mA displays
on console and remote fluoro
control.
9.
Press ACCUM button on console.
ACC indicator will light/extinguish
on console as the switch is
toggled.
10.
Press ACCUM button on remote fluoro control
if used.
ACC indicator will light/extinguish
on remote fluoro control as the
switch is toggled.
11.
Press the pulse fluoro button 18 on the
console (if pulse fluoro option is fitted).
The pulse fluoro indicator will
light/extinguish as the switch is
toggled.
12.
Press power OFF button on console.
Unit switches off.
4.4.3 Generator
Preliminary
Tests
WARNING:
USE EXTREME CAUTION WHEN MEASURING HIGH VOLTAGES.
NOTE: VERIFY THE POSITION OF EACH OF THE SWITCHES OF SW1 ON THE
GENERATOR CPU BOARD FOR THE GENERATOR UNDER TEST AS PER THE
TABLE BELOW. THESE SWITCHES MUST BE SET CORRECTLY PRIOR TO
CONTINUING. REFER TO THE PRODUCT DESCRIPTION (SECTION 1D) AS
NECESSARY FOR PRODUCT DEFINITION.
For GENERATOR POWER, refer to copy of the product description (PD)
form in Chapter 1D
GENERATOR
POWER
SW1-3 SW1-2 SW1-1
30 kW (350 Series)
ON
ON
OFF
37.5 kW (Indico 100 only) ON OFF
ON
50 kW (650 Series)
OFF
ON
ON
65
kW
(850
Series)
OFF ON
OFF
80 kW (1050 Seies)
OFF
OFF
ON
SW1-4: OFF for two filament boards, ON for one filament board.
SW1-5: OFF for 150 max kV, ON for 125 max kV. Refer to PD near top of page 2.
SW1-6: OFF for dual speed, ON for low speed. Refer to PD near top of page 3.
SW1-7: OFF for 2 tube, ON for 1 tube. Refer to PD near top of page 3.
SW1-8: ON. It is strongly suggested this switch not inadvertently be set to OFF!
Setting to OFF resets to factory defaults (calibration, configuration, etc)