3C
Programming the Generator
CPI Canada Inc.
Page 3C-8
Rev. B Millenia / Indico 100 Series Service Manual Ch # 740904-05
3C.5.1 Tube Selection (cont)
* TUBE 1 SELECTION *
DX93HS 0.6/1.5
RO1750 0.6/1.3
DX101HS 0.6/1.3
MX75 1.0/2.0
MX100 0.6/1.25
<< >>
This menu is only available if additional tubes have been downloaded via a computer running CPI
GenWare
software.
* TUBE 1 SELECTION *
***USER DOWNLOADED TUBES DISPLAYED HERE**
<< RETURN
Use these steps to access the TUBE SELECTION menus.
Step
Action
Result
1.
From the GEN CONFIGURATION menu
select TUBE SELECTION.
2.
Select TUBE 1 or TUBE 2. This step not
available on one tube generators, go
directly to step 3.
The available tubes to choose from will
display.
3.
Select the X-ray tube type to be assigned
to that tube location. Use the >> and <<
buttons to navigate through the tube
selection menus if your desired tube is not
displayed on the current screen.
This will assign the selected X-ray tube to
the desired tube location.
4.
Once the desired tube has been selected,
parameters for that tube are displayed
showing the default values. DO NOT
adjust the default values at this time.
See next page for further details. Do not
press RETURN or >> until this is
requested in a later step.
5.
Additional tube types may be downloaded
using the CPI GenWare
utility software.
Refer to section 3C.6.0 DATA LINK.
When the desired tube is selected, the default limits are displayed. Please consult the X-ray tube data
sheet(s) before making any changes.
The dual speed starter operates at 60 or 180 Hz (50 or 150 Hz for some tube types) independent
of line frequency. The low speed starter operates at 50 Hz for 50 Hz mains, or 60 Hz for 60 Hz mains.
Therefore for generators equipped with the low speed starter, the 60 Hz tube ratings are automatically
derated for 50 Hz operation if required.
PLEASE DO NOT CHANGE ANY DEFAULTS UNLESS THE IMPACT OF THOSE CHANGES IS
CLEARLY UNDERSTOOD. INITIAL CALIBRATION SHOULD BE PERFORMED USING THE DEFAULT
VALUES.