4
Acceptance Testing
CPI Canada Inc.
Page 4-14
Rev. F Millenia / Indico 100 Series Service Manual Ch # 740904-08
4.5.2 Generator
Fluoro
Tests
(Cont)
Step
Action
Result
Check
4.
Select 70 kVp fluoro.
Select the fluoro mA values shown below:
A: 1.0 mA
B: 2.0 mA
C: 4.0 mA
D: 6.0 mA
Measure the fluoro mA at TP11 and TP14.
Using scaling factors in figure 4-3 verify mA
values per the RESULTS column:
A: 1.0 mA
±
20%.
B: 2.0 mA
±
20%.
C: 4.0 mA
±
15%.
D: 6.0 mA
±
15%.
5.
Press the PPS + and
-
buttons (if pulse fluoro
option is fitted).
Verify that the pulsed fluoro rate
increase and decreases.
6.
Run a sufficiently long fluoro exposure to
accumulate some time on the console fluoro
display, and on the remote fluoro control if
used
Press the ZERO button on the console and on
the remote fluoro control if applicable.
Verify that the accumulated time
is reset to ZERO after pressing
each of the ZERO buttons.