Programming the Generator
CPI Canada Inc.
Page 3C-12
Rev. B Millenia / Indico 100 Series Service Manual Ch # 740904-05
3C.5.3 Receptor Setup
The RECEPTOR SETUP menus allow each of the image receptors to be programmed as defined in the
table following the example menu screens.
* RECEPTOR SETUP [sym] *
TUBE: 1
AEC CHANNEL: 1
TOMO: NO
INTERFACE OPTS: 0
FLUORO: NO
+
SERIAL: NO
-
EXIT >>
* RECEPTOR SETUP [sym]*
RECEPTOR SYM: [sym]
MEMORY: DEF
FLUORO HANG: 30 SEC
REM TOMO BUT: 2000 MS
RAD HANG: 0 SEC
+
LAST IMAGE HOLD: 40 MS
-
<< >>
* RECEPTOR SETUP [sym]*
SF/LF SWITCH: MAN
DEFAULTS
AEC BACKUP: FIXED
AEC BACKUP MAS: 500
+
AEC BACKUP MS: 3200
-
<<
THE DEFAULTS SELECTION IN THE MENU ABOVE IS ONLY AVAILABLE IF MEMORY IN MENU 2
WAS SET TO DEF.
RECEPTOR MENUS 4 AND 5 BELOW ARE ONLY ACCESSIBLE IF DEFAULTS IS ENABLED.
* RECEPTOR SETUP [sym] DEFAULTS*
TECHNIQUE: AEC
LEFT FIELD: YES
FOCUS: SMALL
CENTER FIELD: YES
FILM SCREEN: 1
RIGHT FIELD: YES
<< >>
* RECEPTOR SETUP [sym] DEFAULTS*
KV:
75
DENSITY: 0
MA: 320
MS:
50
+
-
<<